UC Flow Cytometry Core
William Miller PhD
The Flow Core supports a Beckman CytoFLEX S with 9 detectors and 4 lasers in CARE 2874 and a FACSCalibur 4 color flow cytometer system in CARE 2876. Access is available 24 hrs a day and 7 days per week once investigators are registered in Stratocore and approved by the core director.
To book equipment and/or access services from the Flow Core, please login or create an account in Stratocore at https://ppms.us/uc/start/.
For access to the instruments and the online calendar, please contact Dr. Miller.
-
FACSCalibur System access (2 Laser, 4 color)
The FACSCalibur is configured as follows:
- FL1 Channel - Blue Laser (488nm) and 515-545nm Emission Filter (FITC, GFP, etc.)
- FL2 Channel - Blue Laser (488nm) and 564-601nm Emission Filter (PE, PI, RFP, etc.)
- FL3 Channel - Blue Laser (488nm) and 670LP Emission Filter (7-AAD, PerCP, etc)
- FL4 Channel - Red Laser (635nm) and 653-669nm Emission Filter (APC, Cy5, etc.)
Cost:Request Information$20 per half hour
-
CytoFLEX S System Access (4 laser, 9 color)
Here are the channels currently configured (in bold) the ones in italics can be utilized by repositioning the filters, which is relatively easy to do.
CytoFLEX (4 laser, 9 color)
Violet Laser (405nm) Emmision Filter
PB450 (PacBlue, eFluor450) 450/45
KO525 (Krome Orange, Zombie Aqua) 525/40
BV605 610/20
BV650 660/20
BV786 780/60
Blue Laser (488nm)
FITC (GFP, Alexa488, Venus) 525/40
PerCP (PC5, PC5.5) 690/50
Yellow/Green Laser (561nm)
PE (PI, Sytox Orange, DsRed, TdTomato) 585/42
ECD (mCherry, Texas Red) 610/20
PC7 780/60
PC5 690/50
Red Laser (638nm)
APC (AlexaFluor 647, eFluor660) 660/10
APC-A750 (APC-Cy7, APC eFluor780 780/60
APC-A700 (AlexaFluor 700) &nCost:Request Information$20 per half hour
-
FACSCalibur Four Color Flow Cytometer
The FACSCalibur is configured as follows:
- FL1 Channel - Blue Laser (488nm) and 515-545nm Emission Filter (FITC, GFP, etc.)
- FL2 Channel - Blue Laser (488nm) and 564-601nm Emission Filter (PE, PI, RFP, etc.)
- FL3 Channel - Blue Laser (488nm) and 670LP Emission Filter (7-AAD, PerCP, etc)
- FL4 Channel - Red Laser (635nm) and 653-669nm Emission Filter (APC, Cy5, etc.)
Request Information -
Beckman CytoFLEX S V2-B2-Y3-R2 Flow Cytometer (9 Detectors, 4 Lasers)
Here are the channels currently configured (in bold) the ones in italics can be utilized by repositioning the filters, which is relatively easy to do.
Request Information
CytoFLEX (4 laser, 9 color)
Violet Laser (405nm) Emmision Filter
PB450 (PacBlue, eFluor450) 450/45
KO525 (Krome Orange, Zombie Aqua) 525/40
BV605 610/20
BV650 660/20
BV786 780/60
Blue Laser (488nm)
FITC (GFP, Alexa488, Venus) 525/40
PerCP (PC5, PC5.5) 690/50
Yellow/Green Laser (561nm)
PE (PI, Sytox Orange, DsRed, TdTomato) 585/42
ECD (mCherry, Texas Red) 610/20
PC7 780/60
PC5 690/50
Red Laser (638nm)
APC (AlexaFluor 647, eFluor660) 660/10
APC-A750 (APC-Cy7, APC eFluor780 780/60
APC-A700 (AlexaFluor 700)